Electrical Safety Analyzer 19032 Users Manual
4-8
(3) When set to S T O P, if any test
STEPs fail, press STOP, then press
START to restart the test.
AC OFFSET
0 ~ 2.5mA
0.10mA
(1) When Offset value is higher than
AC OFFSET
value, Current reading
= Current real measurement value
Offset value.
(2) When Offset value is lower than
AC
OFFSET value, Current value
=(Real measurement value)
2
(Offset)
2
LC OFFSET
0 ~ 2.5mA
0.00mA
(1) When Offset value is higher than
LC OFFSET
value, Current reading
= Current real measurement value
Offset value.
(2)
When Offset value is lower than LC
OFFSET value, Current reading =
(Real measurement value)
2
(Offset)
2
IEEE-488.1
ON/OFF
OFF
(1) When set to ON, GPIB/RS-232
command format is IEEE-488.1
(compatible with 9032C).
(2) When set to OFF, GPIB/RS-232
command format is IEEE-488.2.
Step Signal
ON/OFF
OFF
(1) When Step Signal is set to ON, the
signals on the rear panel are as
follows:
1. UNDER TEST = EOT
2. PASS = EOS
3. FAIL = PASS/FAIL (when the
signal is PASS, RELAY OPEN.
When the signal is FAIL, RELAY
CLOSE.)
(2) When Step Signal is OFF, the
signals of UNDER TEST, PASS
and
FAIL on rear panel follows the
original definition.
Note
When the Step Signal is ON,
the PASS setting is disabled
since the PASS signal changes
to EOS.
LC OFFS GET
ON/OFF
ON
(1) When set to ON, LC Mode will be
included when performing
OFFSET
GET.
(2) When set to OFF, LC Mode will not
be included when performing
OFFSET GET.