EasyManua.ls Logo

Chroma 19032 - Page 84

Default Icon
138 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Electrical Safety Analyzer 19032 User’s Manual
5-18
[:SOURce]:SAFEty:STEP<n>:GB:CURRent:OFFSet <numeric value>
This command sets the Offset value of the ground bond test for the selected step.
Parameter <numeric value>: Offset value of ground bond test in ohms (Ω).
Range: 0.000 ~ 0.5000
Example: Input command “SAFE:STEP1:GB:CURR:OFFS 0.1”
Description: This command sets the Offset value of the ground bond test in STEP 1
to 100mΩ.
[:SOURce]:SAFEty:STEP<n>:GB:CURR:OFFSet?
This command queries the Offset value of the ground bond test for the selected
step.
Example: Input command “SAFE:STEP1:GB:CURR:OFFS?”
Return message “+1.000000E-01”
Description: Return message “+1.000000E-01” means the Offset value for the
ground bond test in STEP 1 is 100mΩ.
[:SOURce]:SAFEty:STEP<n>:GB[:LEVel] < numeric value >
This command sets the grounding resistance test current value in amperes (A) for
the selected step.
Range : The range can be set as follows:
1. 30:30 : 1 ~30
2. 30:40 : 1 ~40
3. 30:45 : 1 ~45
4. 30:60 : 1 ~60
Example : Input command “SAFE:STEP1:GB 5”
Description : This command sets the step 1 current value to 5A.
[:SOURce]:SAFEty:STEP<n>:GB[:LEVel]?
This command queries the grounding resistance test current value for the selected
step.
Example : Input command “SAFE:STEP:GB?”
Return message “+5.000000E+00”
Description : Return message “+5.000000E+00” means the current value of the
grounding resistance test is 5A.
[:SOURce]:SAFEty:STEP<n>:GB:LIMit[:HIGH] < numeric value >
This command sets the grounding resistance high limit value in Ohms for the
selected step.
Range : 0.0001 ~ 0.51 (the high limit value of grounding resistance x
current value should be ≤ 6.3V)
Example : Input command “SAFE:STEP1:GB:LIM 0.11”
Description : This command sets the step 1 grounding resistance high limit
value to 0.11 ohm.
[:SOURce]:SAFEty:STEP<n>:GB:LIMit[:HIGH]?
This command queries the grounding resistance high limit value for the selected
step.
Example : Input command “SAFE:STEP:GB:LIM?”
Return message “+1.100000E-01”
Description : Return message “+1.100000E-01” means the grounding resistance
high limit value is 0.11 ohm.

Table of Contents