Electrical Safety Analyzer 19032 User’s Manual
5-18
[:SOURce]:SAFEty:STEP<n>:GB:CURRent:OFFSet <numeric value>
This command sets the Offset value of the ground bond test for the selected step.
Parameter <numeric value>: Offset value of ground bond test in ohms (Ω).
Range: 0.000 ~ 0.5000
Example: Input command “SAFE:STEP1:GB:CURR:OFFS 0.1”
Description: This command sets the Offset value of the ground bond test in STEP 1
to 100mΩ.
[:SOURce]:SAFEty:STEP<n>:GB:CURR:OFFSet?
This command queries the Offset value of the ground bond test for the selected
step.
Example: Input command “SAFE:STEP1:GB:CURR:OFFS?”
Return message “+1.000000E-01”
Description: Return message “+1.000000E-01” means the Offset value for the
ground bond test in STEP 1 is 100mΩ.
[:SOURce]:SAFEty:STEP<n>:GB[:LEVel] < numeric value >
This command sets the grounding resistance test current value in amperes (A) for
the selected step.
Range : The range can be set as follows:
1. 30:30 : 1 ~30
2. 30:40 : 1 ~40
3. 30:45 : 1 ~45
4. 30:60 : 1 ~60
Example : Input command “SAFE:STEP1:GB 5”
Description : This command sets the step 1 current value to 5A.
[:SOURce]:SAFEty:STEP<n>:GB[:LEVel]?
This command queries the grounding resistance test current value for the selected
step.
Example : Input command “SAFE:STEP:GB?”
Return message “+5.000000E+00”
Description : Return message “+5.000000E+00” means the current value of the
grounding resistance test is 5A.
[:SOURce]:SAFEty:STEP<n>:GB:LIMit[:HIGH] < numeric value >
This command sets the grounding resistance high limit value in Ohms for the
selected step.
Range : 0.0001 ~ 0.51 (the high limit value of grounding resistance x
current value should be ≤ 6.3V)
Example : Input command “SAFE:STEP1:GB:LIM 0.11”
Description : This command sets the step 1 grounding resistance high limit
value to 0.11 ohm.
[:SOURce]:SAFEty:STEP<n>:GB:LIMit[:HIGH]?
This command queries the grounding resistance high limit value for the selected
step.
Example : Input command “SAFE:STEP:GB:LIM?”
Return message “+1.100000E-01”
Description : Return message “+1.100000E-01” means the grounding resistance
high limit value is 0.11 ohm.