EasyManua.ls Logo

Chroma 19032 - Page 94

Default Icon
138 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Electrical Safety Analyzer 19032 User’s Manual
5-28
[:SOURce]:SAFEty:STEP<n>:IR:TIME:RAMP < numeric value >
This command sets the ramp time in seconds (s) for the selected step.
Range: 0 or 0.1~999. (0=OFF)
Example: Input command “SAFE: STEP4: IR: TIME: RAMP 0.5”
Description: This command sets the ramp time for STEP 4 to 0.5 sec.
[:SOURce]:SAFEty:STEP<n>:IR:TIME:RAMP?
This command queries the ramp time for the selected step.
Example: Input command “SAFE: STEP4: IR: TIME: RAMP?”
Return message “5.000000E-01”
Description: Return message “5.000000E-01” means the ramp time for STEP 4 is
set to 0.5 sec.
[:SOURce]:SAFEty:STEP<n>:IR:TIME[:TEST] < numeric value >
This command sets the test time in seconds (s) for the selected step.
Range: 0 or 0.3~999. (0=CONTINUOUS)
Example: Input command “SAFE:STEP4:IR:TIME 1”
Description: This command sets the test time for STEP 4 to 1.0sec.
[:SOURce]:SAFEty:STEP<n>:IR:TIME[:TEST]?
This command queries the test time for the selected step.
Example: Input command “SAFE:STEP4:IR:TIME?”
Return message “1.000000E+00”
Description: Return message “1.000000E+00” means the test time for STEP 4 is
set to 1 sec.
[:SOURce]:SAFEty:STEP<n>:IR:TIME:FALL < numeric value >
This commands sets the time required for the voltage to fall to 0 in seconds (s) for
the selected step.
Range: 0 or 0.1~999.0. (0=OFF)
Example: Input command “SAFE:STEP4:IR:TIME:FALL 3”
Description: This command sets the time required for the voltage to fall to 0 for
STEP 3 to 3.0sec
[:SOURce]:SAFEty:STEP<n>:IR:TIME:FALL?
This command queries the time required for the voltage to fall to 0 for the selected
step.
Example: Input command “SAFE:STEP4:IR:TIME:FALL?”
Return message “3.000000E+00”
Description: Return message “3.000000E+00” means the time required for the
voltage to fall to 0 for STEP 3 is set to 3.0sec.
[:SOURce]:SAFEty:STEP<n>:IR:RANGe[:UPPer]
This command selects the range upper than the current measured according to
current value
users inputted. The unit is ampere (A).
Range: 0~0.01
Example: Input command “SAFE:STEP4:IR:RANG 0.0003”
Description: This command sets IR measured current value of main STEP 4 to
300uA thus the selected IR range upper than the current measured is
3mA.

Table of Contents