GPIB/RS232 Operation Description (IEEE-488.2)
5-31
[: SOURce]:SAFEty:STEP<n>:PAuse:TIME[:TEST] <numeric_value>
This command sets the time required of PA mode test for the selected step.
Example: Input command “SAFE:STEP5:PA:TIME 5”
Description: This command sets the time required for STEP 5 to 5.0sec.
[: SOURce]:SAFEty:STEP<n>:PAuse:TIME[:TEST]?
This command queries the time required of PA mode test for selected step.
Example: Input command “SAFE:STEP5:PA:TIME ?”
Return message “5.000000E+00”
Description: Return message “5.000000E+00” means the test time required for
STEP 5 is set to 5.0sec.
[:SOURce]:SAFEty:STEP<n>:OSC:CRANge? <MAXimun|MINimum|NOW>
This command queries the maximum, minimum, and current operating ranges of
the open/short circuit detection mode for the selected step.
Example: Input command “SAFE:STEP6:OSC:CRAN? MAX”.
Return message “3”
Description: Return message “3” means the maximum range that can be set in
Step 6 is 3.
[:SOURce]:SAFEty:STEP<n>:OSC:CURRent<m>:OFFSet <numeric value>
This command sets the Offset value range of the open/short detection mode for the
selected step.
Parameter m: The range of open/short detection mode.
Parameter <numeric value>: The Offset value of open/short circuit detection
mode. The unit is in Farad (F).
Range: Range 1, 2 and 3 are all 0~0.0000009999.
Example: Input command “SAFE:STEP6:OSC:CURR3:OFFS 0.00000001”.
Description: This command sets the Offset value of range 3 of the open/short
circuit detection mode in Step 6 to 10nF.
[:SOURce]:SAFEty:STEP<n>:OSC:CURRent<m>:OFFSet?
This command queries the Offset value range of the open/short circuit detection
mode for the selected step.
Parameter m: The range of open/short detection mode.
Example: Input command “SAFE:STEP6:OSC:CURR3:OFFS?”.
Return message “+1.000000E-08”.
Description: Return message “+1.000000E-08” means the Offset value under
open/short detection mode in Step 6 is 10nF.
[:SOURce]: SAFEty: STEP<n>: OSC: LIMit[:OPEN] < numeric value >
This command sets the open circuit percentage in percentage (100%) for the
selected step
Range: 0.1~1.0
Example: Input command “SAFE:STEP6:OSC:LIM:OPEN 0.3”
Description: This command sets the open circuit percentage for STEP 6 to 30%.
[:SOURce]: SAFEty: STEP<n>: OSC: LIMit[:OPEN]?
This command queries the open circuit percentage for the selected step.
Example: Input command “SAFE:STEP6:OSC:LIM:OPEN?”
Return message “3.000000E-01”
Description: Return message “3.000000E-01” means the open circuit percentage
for STEP 6 is set to 30%.