GPIB/RS232 Operation Description (IEEE-488.2)
5-33
mode in Farads (F).
Cs range: Range 1: 0.001~9.999nF
Range 2: 0.01~99.99nF
Range 3: 0.1~500.0nF
Example: Input command “SAFE:STEP6:OSC:CSTandard 1, 0.000000001”
Description: This command sets the Cs value in Range 1 under the open/short
circuit mode in Step 6 to 1nF.
[:SOURce]:SAFEty:STEP<n>:OSC:CSTandard?
This command queries the Cs value for the selected step.
Example: Input command “SAFE:STEP6:OSC:CSTandard?”
Return message “+1.000000E-09”
Description: Return message “+1.000000E-09” means the Cs value for Step 6 is
set to 1nF.
[:SOURce]:SAFEty:STEP<n>:LC:CURRent:OFFSet[:LC] <numeric value>
This command sets the Offset value for the human body impedance model for the
selected step.
Parameter <numeric value>: The Offset value of human body simulation circuit
Test in Amperes (A). The Offset value must be
greater than 0.
6000-05/07 range:
When the High Limit setting range is from 0.1uA to 59.9uA, the
High Limit + Offset setting value ≤ 0.000066.
When the High Limit setting range is from 60uA to 599uA, the
High Limit + Offset setting value ≤ 0.00066.
When the High Limit setting is ≥ 600uA:
High Limit + Offset setting value of UL544NP ≤ 0.0066.
High Limit + Offset setting value of UL544P ≤ 0.011.
High Limit + Offset setting value of UL1563 ≤ 0.011.
High Limit + Offset setting value of UL2601 ≤ 0.011.
High Limit + Offset setting value of UL1950 ≤ 0.011.
6000-04/06/08 range:
When the High Limit setting range is from 0.1uA to 599uA, the
High Limit + Offset setting value ≤ 0.00066.
When the High Limit setting is ≥ 600uA:
High Limit + Offset setting value of UL544NP ≤ 0.0066.
High Limit + Offset setting value of UL544P ≤ 0.011.
High Limit + Offset setting value of UL1563 ≤ 0.011.
High Limit + Offset setting value of UL2601 ≤ 0.011.
High Limit + Offset setting value of UL1950 ≤ 0.011.
High Limit + Offset setting value of 1950-U1 (RMS) ≤ 0.055.
High Limit + Offset setting value of 1950-U1 (PEAK) ≤ 0.077.
High Limit + Offset setting value of 2601-U1 ≤ 0.011.
Note: Only 6000-08 can set DEVICE 1950-U1 and 2601-U1.
Example: Input command “SAFE:STEP7:LC:CURR:OFFS 0.000001”.
Description: This command sets the Offset value of the human body impedance
model in Step 7 to 10uA.