Calibration Procedure
8. Calibration Procedure
Before process this section the HI-POT tester be warm up at least 30 minutes. Take off the
calibration front panel. Press the lock switch, this is a hardware data backup protection
circuit, to avoid calibration data loss.
The following items are need to calibration.
Voltage Calibration (See 8.2)
ACV 5kV Offset (100V) ; AC Voltage OFFSET point
ACV 5kV Full (4kV) ; AC Voltage FULL point
DCV 6kV Offset (100V) ; DC Voltage OFFSET point
DCV 6kV Full (4kV) ; DC Voltage FULL point
IRV 1kV Offset (100V) ; IR Voltage OFFSET point
IRV 1kV Full (1kV) ; IR Voltage FULL point
Current Calibration (See 8.3)
ACA 3mA Offset (0.12mA) ; AC total current 2.99mA range OFFSET point
ACA 3mA Full (2.5mA) ; AC total current 2.99mA range FULL point
ACA30mA Offset (2.5mA) ; AC total current 30mA range OFFSET point
ACA 30mA Full (25mA) ; AC total current 30mA range FULL point
RACA 3mA Offset (0.12mA) ; AC real current 2.99mA range OFFSET point
RACA 3mA Full (2.5mA) ; AC real current 2.99mA range FULL point
RACA 30mA Offset (2.5mA) ; AC real current 30mA range OFFSET point
RACA 30mA Full (25mA) ; AC real current 30mA range FULL point
DCA 3mA Offset (0.12mA) ; DC 2.99mA range OFFSET point
DCA 3mA Full (2.5mA) ; DC 2.99mA range FULL point
DCA 10mA Offset (2.5mA) ; DC 10mA range OFFSET point
DCA 10mA Full (8mA) ; DC 10mA range FULL point
Withstanding Voltage Mode ARCing Calibration (See 8.4)
AC ARC 15mA (5mA) ; AC ARCing calibration
DC ARC 10mA (5mA) ; DC ARCing calibration
Insulation Resistance Mode Leakage Current Meter Calibration (See 8.5)
IRR Range1 (1GΩ) ; IR Resistor range 1 Calibration
IRR Range2 (100MΩ) ; IR Resistor range 2 Calibration
IRR Range3 (10MΩ) ; IR Resistor range 3 Calibration
IRR Range4 (10MΩ) ; IR Resistor range 4 Calibration
Ground Continue Calibration (See 8.6)
Contrast Calibration (See 8.7)