HIPOT Tester 19051/19052/19053/19054 User’s Manual
8.9.2 Commands Summary
:CALibration:STATe <Boolean>
The command is used to select if the calibration data applying (1) or (0).
At *RST, the state is set to ON.
:CALibration:REQuest?
The event attempts to attain the calibrating this device and returns 1 if succeed and 0 if it
fails.
:CALibration:VALue <Numeric_Value>
Enters the value of calibration. If the state is not calibration or output on, an error –203 will
be generated in addition to the execution error.
:CALibration:SAFEty:STARt
Start output the source, and can enter the value of calibration. When the state is not
calibration, an error –203 will be generated as executing this command.
:CALibration:SAFEty:STOP
Stop output the source.
:CALibration:SAFEty:AC:VRANge? {MAXimum, MINimum }
Query the maximum and minimum of the range of the voltage source and meter of the AC
mode.
:CALibration:SAFEty:AC:VOLTage (range):OFFSet[:SOURce] [:VOLTage]
[<Numeric Value>]
Changing to the offset item of the voltage source and voltage meter of AC mode,
and set the
output voltage value that is used in the calibration.
:CALibration:SAFEty:AC:VOLTage (range):OFFSet[:SOURce] [:VOLTage] ?
Change to the offset item of the voltage source and voltage meter of AC mode, and returns
the output voltage value which is used in the calibration.
:CALibration:SAFEty:AC:VOLTage (range):OFFSet:BEST?
Change to the offset item of voltage source and voltage meter of AC mode, and returns the
best value that is selected by device.
:CALibration:SAFEty:AC:VOLTage (range):FULL[:SOURce] [:VOLTage]
[<Numeric Value>]
Change to the offset item of voltage source and voltage meter of AC mode, and set the
output voltage value that is used in the calibration.
:CALibration:SAFEty:AC:VOLTage (range):FULL[:SOURce] [:VOLTage] ?
Change to the offset item of voltage source and voltage meter of AC mode, and returns the
output voltage value which is used in the calibration.
:CALibration:SAFEty:AC:VOLTage (range):FULL:BEST?
Change to the offset item of voltage source and voltage meter of AC mode, and returns the
best value that is selected by device.
:CALibration:SAFEty:AC:CRANge? { MAXimum, MINimum }
8-10