HIPOT Tester 19051/19052/19053/19054 User’s Manual
3. Press [UP] or [DOWN] until LCD contrast brightness is appropriate.
8.8 Finish Calibration
press [EXIT] [DOWN] [DOWN] [DOWN] [ENTER]
display PASSWORD:
press [A] [A] [A] [A] [ENTER]
display CALIBRATION IS OFF or CALIBRATION IS ON; choose
CALIBRATION IS ON, if shows CALIBRATION IS OFF.
press [A] [A] [A] [A] [ENTER]
display CALIBRATION IS ON
press [EXIT] to complete calibration steps.
8.9 Remote Calibration Command
8.9.1 Command List
CALibration
:STATe <Boolean>
:REQuest?
:VALue <Numeric Value>
:SAFEty
:STARt
:STOP
:AC
:VRANge? MAXimum|MINimum
:VOLTage ( range)
:OFFSet
[:SOURce]
[:VOLTage] [<Numeric Value>]
[:VOLTage] ?
:BEST?
:FULL
[:SOURce]
[:VOLTage] [<Numeric Value>]
[:VOLTage] ?
:BEST?
:CRANge? MAXimum|MINimum
:CURRent ( range)
:OFFSet
[:SOURce]
[:VOLTage] [<Numeric Value>]
[:VOLTage] ?
:BEST?
:FULL
[:SOURce]
[:VOLTage] [<Numeric Value>]
[:VOLTage] ?
:BEST?
:RCRANge? MAXimum|MINimum
:RCURrent ( range)
:OFFSet
[:SOURce]
[:VOLTage] [<Numeric Value>]
[:VOLTage] ?
:BEST?
:FULL
[:SOURce]
8-8