GPIB Operation Description (Option)
5.6 IEEE 488.2 Command
* CLS command
Clear status data configuration the following actions are needed.
Clear standard event status register.
Clear status bit group register except for MAV bit (bit 4).
* ESE <metric system value> Command
Use setting standard event status enable register value, <metric system value> range is 0
∼255.
* ESE? Command
The controller is used for query standard event status of device enable register value.
The output format is <metric system value>, its’ range is 0 ∼255.
* ESR? Command
The controller queries the standard event register value of the device. After performing
this command, the standard event register value will be cleared to 0.
The output format is <metric system value>, its’ range is 0 ∼ 255.
* SRE <metric system value> Command
Use for setting service request register value, its’ <metric system value> value is 0 ∼ 255.
* SRE? Command
The controller is reading service request enable register initial setting.
The output format is <metric system value>, its’ range is 0 ∼255.
* STB? Command
The controller is for reading status bit register value.
The output format is <metric system value>, its’ range is 0 ∼255.
* OPC Command
Operation is completed command
* OPC? Command
Operation complete query command.
The output format is ASCII character " 1 ".
* PSC 0|1 Command
Power on status clear command.
* PSC? Command
Power on status clear query command.
The output format is ASCII character " 1 " or " 0 ".
* RST Command
The device reset command.
* IDN? Command
The controller is for reading the basic data of the device.
The output format separate four fields by comma, it denote separately: manufacturer,
device model, serial number and firmware version.
5-5