Specifications
2. Specifications (18°C ∼ 28°C RH ≤ 70%)
Scan Unit
8 ports,
⋅ ±phase (19053 only), 4 ports, ⋅ ±phase (19054 only)
Withstanding Voltage Test
Test Voltage
AC: 0.05 ∼ 5kV/ DC: 0.05 ∼ 6kV Constant Voltage
Voltage Regulation
≤ (1%+5V), Rated Load
V-display Accuracy ± (1% of reading + 5 counts), 2V resolution
Cutoff Current
(Note 2)
AC: 0.1mA ~ 30mA (Note 1), DC: 0.01mA ~ 10mA (Note 1),
0.1uAdc resolution
Current Accuracy
(Note2)
± (1% of reading + 5 counts)
Real Current ± (5% of total current + 20 counts) (Note2) WAC
only
Current Display
Hi limit setting Display Range
< 300uA: 0.1uA~299.9uA (dc only)
< 3mA: 0.001mA~2.999mA
<30mAac (10mAdc): 0.01mA~30.00mAac (10mAdc)
Output Frequency 50Hz, 60Hz
Test Time (Note 3) 0.3 ~ 999 Sec, continue (Note 1) (0.2S for LCD off)
Ramp Time 0.1 ~ 999 Sec, off (Note 1)
Fall Time
0.1 ∼ 999 Sec, off
Judgment Delay
(Wdc Only)
0.1 ~ 99.9 Sec. (Note 1)
Arc Detection (Note 4)
Setting Mode Programmable Setting
Detection Current AC: 1mA ~ 15mA, DC: 1mA ~ 10mA
Min. pulse width 10us approx.
GOOD/NO-GO Judgment Function
Judgment System
• Window comparator.
• A NO-GO judgment is made when a current greater than
the high limit value or smaller than the low limit value is
detected.
• When a NO-GO judgment is made, the output voltage is
cut out and a NO-GO alarm signal is delivered.
• If no abnormal state is detected during the test time a
GOOD judgment is made and a GOOD signal is delivered.
Insulation Resistance Test (19052, 19053, 19054 only)
Test Voltage DC: 0.05kV ~ 1kV, Constant Voltage
V-display Accuracy ± (1.5% of reading + 5 counts) (open voltage), 2V resolution
Resistance Range
0.1 MΩ ∼ 10 GΩ (19052 up to 50GΩ)
Measuring Accuracy
≥ 500: 1 MΩ ∼ 1GΩ: ± (5% of reading + 5 counts)
1GΩ ∼ 10 GΩ: ± (10% of reading + 5 counts)
10GΩ ∼50 GΩ: ± (15% of reading + 5 counts) (19052 only)
< 500V: 0.1 MΩ ∼ 1GΩ: ± (10% of reading + 5 counts)
2-1