HIPOT Tester 19051/19052/19053/19054 User’s Manual
5.8 Error Messages
y Error messages are saved in error queue which accessed by FIFO method. The
return first error message is the first being saved.
y When the error message is over 30, the last position will save -350," Queue
overflow". The error queue can’t save error message any more till there is error
message out.
y When there is no error, the first position will save +0,"No error" in error queue.
-102 Syntax error
Syntax error, usually includes not allowed character symbol in command.
-103 Invalid separator
Invalid separator found in command string
-108 Parameter not allowed
The device receives parameter is not allowed.
-109 Missing parameter
Parameter is missed.
-112 Program mnemonic too long
Simple command program header is over 12 characters.
-113 Undefined header
The device is received undefined header.
-114 Header suffix out of range
Variable is out of range.
-120 Numeric data error
Numerical parameter error
-140 Character data error
Inputted character data error
-151 Invalid string data
Invalid string data is usually missing double quotation.
-158 String data not allowed
The device is received disallowed string data.
-170 Expression error
The device is received uncompleted parameter data, such as missing the right
parenthesis.
-200 Execution error
Command executing error
-203 Command protected
The device does not receive this command.
-221 Settings conflict
The device does not receive this command.
-222 Data out of range
The data is out of range.
-223 Too much data
Received string length is over, can’t execute.
-290 Memory use error
Store or read the memory is error.
-291 Out of memory
The value is out of memory.