Calibration Procedure
Query the maximum and minimum of the range of the current meter of the AC mode.
:CALibration:SAFEty:AC:CURRent (range):OFFSet[:SOURce] [:VOLTage]
[<Numeric Value>]
Change to the offset item of current meter of AC mode, and set the output voltage value that
is used in the calibration. The header suffix (range) is selected the range of current.
:CALibration:SAFEty:AC:CURRent (range):OFFSet[:SOURce] [:VOLTage] ?
Change to the offset item of current meter of AC mode, and returns the output voltage value
which is used in the calibration.
:CALibration:SAFEty:AC:CURRent (range):OFFSet:BEST?
Change to the offset item of current meter of AC mode, and returns the best value that is
selected by device.
:CALibration:SAFEty:AC:CURRent (range):FULL[:SOURce] [:VOLTage]
[<Numeric Value>]
Change to the full item of current meter of AC mode, and set the output voltage value that is
used in the calibration.
:CALibration:SAFEty:AC:CURRent (range):FULL[:SOURce] [:VOLTage] ?
Change to the full item of current meter of AC mode, and returns the output voltage value
which is used in the calibration.
:CALibration:SAFEty:AC:CURRent (range):FULL:BEST?
Change to the full item of current meter of AC mode, and returns the best value that is
selected by device.
:CALibration:SAFEty:AC:RCRAnge? { MAXimum, MINimum }
Query the maximum and minimum of the range of the real current meter of the AC mode.
:CALibration:SAFEty:AC:RCURrent (range):OFFSet[:SOURce] [:VOLTage]
[<Numeric Value>]
Change to the offset item of the real current meter of AC mode, and set the output voltage
value that is used in the calibration. The header suffix (range) is selected the range of
current
:CALibration:SAFEty:AC:RCURrent (range):OFFSet[:SOURce] [:VOLTage] ?
Change to the offset item of the real current meter of AC mode, and returns the output
voltage value which is used in the calibration.
:CALibration:SAFEty:AC:RCURrent (range):OFFSet:BEST?
Change to the offset item of real current meter of AC mode, and returns the best value that is
selected by device.
:CALibration:SAFEty:AC:RCURrent (range):FULL[:SOURce] [:VOLTage]
[<Numeric Value>]
Change to the full item of the real current meter of AC mode, and set the output voltage value
that is used in the calibration.
:CALibration:SAFEty:AC:RCURrent (range):FULL[:SOURce] [:VOLTage] ?
Change to the full item of the real current meter of AC mode, and returns the output voltage
value which is used in the calibration.