HIPOT Tester 19051/19052/19053/19054 User’s Manual
:CALibration:SAFEty:AC:RCURrent (range):FULL:BEST?
Change to the full item of real current meter of AC mode, and returns the best value that is
selected by device.
:CALibration:SAFEty:AC:ARANge? {MAXimum, MINimum }
Query the maximum and minimum of the range of the arc meter of the AC mode.
:CALibration:SAFEty:AC:ARC (range):SLOPe[:SOURce] [:VOLTage] [<Numeric
Value>]
Change to the arc item of AC mode, and set the output voltage value that is used in the
calibration.
:CALibration:SAFEty:AC:ARC (range):SLOPe[:SOURce] [:VOLTage] ?
Change to the arc item of AC mode, and returns the output voltage value which is used in the
calibration.
:CALibration:SAFEty:AC:ARC (range):SLOPe:LEVel [<Numeric Value>]
Change to the arc item of AC mode, and set the arc limit value.
:CALibration:SAFEty:AC:ARC (range):SLOPe:LEVel?
Change to the arc item of AC mode, and returns the arc limit value.
:CALibration:SAFEty:AC:ARC (range):SLOPe:BEST?
Change to the arc item of AC mode, and returns the best arc limit value that is selected by
device.
:CALibration:SAFEty:DC:VRANge? {MAXimum, MINimum}
Query the maximum and minimum of the range of the voltage source and meter of the DC
mode.
:CALibration:SAFEty:DC:VOLTage (range):OFFSet[:SOURce] [:VOLTage]
[<Numeric Value>]
Change to the offset item of the voltage source and voltage meter of DC mode, and set the
output voltage value that is used in the calibration.
:CALibration:SAFEty:DC:VOLTage (range):OFFSet[:SOURce] [:VOLTage] ?
Change to the offset item of the voltage source and voltage meter of DC mode, and returns
the output voltage value which is used in the calibration.
:CALibration:SAFEty:DC:VOLTage (range):OFFSet:BEST?
Change to the offset item of voltage source and voltage meter of DC mode, and returns the
best value that is selected by device.
:CALibration:SAFEty:DC:VOLTage (range):FULL[:SOURce] [:VOLTage]
[<Numeric Value>]
Change to the offset item of voltage source and voltage meter of DC mode, and set the
output voltage value that is used in the calibration.
:CALibration:SAFEty:DC:VOLTage (range):FULL[:SOURce] [:VOLTage] ?
Change to the offset item of voltage source and voltage meter of DC mode, and returns the
output voltage value which is used in the calibration.
:CALibration:SAFEty:DC:VOLTage (range):FULL:BEST?
Change to the offset item of voltage source and voltage meter of DC mode, and returns the
8-12