Calibration Procedure
best value that is selected by device.
:CALibration:SAFEty:DC:CRANge? { MAXimum, MINimum }
Query the maximum and minimum of the range of the current meter of the DC mode.
:CALibration:SAFEty:DC:CURRent (range):OFFSet[:SOURce] [:VOLTage]
[<Numeric Value>]
Change to the offset item of current meter of DC mode, and set the output voltage value that
is used in the calibration. The header suffix (range) is selected the range of current.
:CALibration:SAFEty:DC:CURRent (range):OFFSet[:SOURce] [:VOLTage] ?
Change to the offset item of current meter of DC mode, and returns the output voltage value
which is used in the calibration.
:CALibration:SAFEty:DC:CURRent (range):OFFSet:BEST?
Change to the offset item of current meter of DC mode, and returns the best value that is
selected by device.
:CALibration:SAFEty:DC:CURRent (range):FULL[:SOURce] [:VOLTage]
[<Numeric Value>]
Change to the full item of current meter of DC mode, and set the output voltage value that is
used in the calibration.
:CALibration:SAFEty:DC:CURRent (range):FULL[:SOURce] [:VOLTage] ?
Change to the full item of current meter of DC mode, and returns the output voltage value
which is used in the calibration.
:CALibration:SAFEty:DC:CURRent (range):FULL:BEST?
Change to the full item of current meter of DC mode, and returns the best value that is
selected by device.
:CALibration:SAFEty:DC:ARANge? {MAXimum, MINimum }
Query the maximum and minimum of the range of the arc meter of the DC mode.
:CALibration:SAFEty:DC:ARC (range):SLOPe[:SOURce] [:VOLTage] [<Numeric
Value>]
Change to the arc item of DC mode, and set the output voltage value that is used in the
calibration.
:CALibration:SAFEty:DC:ARC (range):SLOPe[:SOURce] [:VOLTage] ?
Change to the arc item of DC mode, and returns the output voltage value which is used in
the calibration.
:CALibration:SAFEty:DC:ARC (range):SLOPe:LEVel [<Numeric Value>]
Change to the arc item of DC mode, and set the arc limit value.
:CALibration:SAFEty:DC:ARC (range):SLOPe:LEVel?
Change to the arc item of DC mode, and returns the arc limit value.
:CALibration:SAFEty:DC:ARC (range):SLOPe:BEST?
Changing to the arc item of DC mode, and returns the best arc limit value that is selected by
device.
:CALibration:SAFEty:IR:VRANge? {MAXimum, MINimum }
Query the maximum and minimum of the range of the voltage source and meter of the IR
8-13