HIPOT Tester 19051/19052/19053/19054 User’s Manual
8-14
mode.
:CALibration:SAFEty:IR:VOLTage (range):OFFSet[:SOURce] [:VOLTage]
[<Numeric Value>]
Change to the offset item of the voltage source and voltage meter of IR mode, and set the
output voltage value that is used in the calibration.
:CALibration:SAFEty:IR:VOLTage (range):OFFSet[:SOURce] [:VOLTage] ?
Change to the offset item of the voltage source and voltage meter of IR mode, and returns
the output voltage value which is used in the calibration.
:CALibration:SAFEty:IR:VOLTage (range):OFFSet:BEST?
Change to the offset item of voltage source and voltage meter of IR mode, and returns the
best value that is selected by device.
:CALibration:SAFEty:IR:VOLTage (range):FULL[:SOURce] [:VOLTage] [<Numeric
Value>]
Change to the full item of voltage source and voltage meter of IR mode, and set the output
voltage value that is used in the calibration.
:CALibration:SAFEty:IR:VOLTage (range):FULL[:SOURce] [:VOLTage] ?
Change to the full item of voltage source and voltage meter of IR mode, and returns the
output voltage value which is used in the calibration.
:CALibration:SAFEty:IR:VOLTage (range):FULL:BEST?
Change to the full item of voltage source and voltage meter of IR mode, and returns the best
value that is selected by device.
:CALibration:SAFEty:IR:RRANge? { MAXimum, MINimum }
Query the maximum and minimum of the range of the resistance meter of the IR mode.
:CALibration:SAFEty:IR:RESistance (range):SLOPe[:SOURce] [:VOLTage]
[<Numeric Value>]
Change to the resistance item of IR mode, and set the output voltage value that is used in
the calibration.
:CALibration:SAFEty:IR:RESistance (range):SLOPe[:SOURce] [:VOLTage] ?
Change to the resistance item of IR mode, and returns the output voltage value which is used
in the calibration.
:CALibration:SAFEty:IR:RESistance (range):SLOPe:BEST?
Change to the resistance item of IR mode, and returns the best value that is selected by
device.
www.valuetronics.com