HIPOT Tester 19051/19052/19053/19054 User’s Manual
: SOURce: SAFEty: STEP: IR: TIME: RAMP?
This command queries selected STEP which test ramp to setting
voltage needed time.
Example: Input command “SAFE: STEP 3: IR: TIME: RAMP?”
Return message “5.000000E-01”
Description: Return message “5.000000E-01” means test ramp to
setting voltage needed time of the main unit STEP 3 is
0.5 sec.
[: TEST] <numeric_value>
: SOURce: SAFEty: STEP: IR: TIME: TEST
This command sets selected STEP which test needed time.
Example: Input command “SAFE:STEP 3:IR:TIME 1”
Description: This command means test needed time of the main unit
STEP 3 is set to 1.0 sec.
[: TEST]?
: SOURce: SAFEty: STEP: IR: TIME: TEST?
This command queries selected STEP which test needed time.
Example: Input command “SAFE:STEP 3:IR:TIME?”
Return message “1.000000E+00”
Description: Return message “1.000000E+00” means test needed
time of the main unit STEP 3 is 1 sec.
: FALL <numeric
value>
: SOURce: SAFEty: STEP: AC: TIME: FALL
This command sets selected STEP which setting voltage value of
falling to zero needed time.
Example: Input command “SAFE: STEP 3: IR: TIME: FALL 0.3”
Description: This command means setting voltage value of the main
unit STEP 3 fall to zero needed time is set to 0.3 sec.
: FALL?
: SOURce: SAFEty: STEP: AC: TIME: FALL?
This command queries selected STEP which setting voltage value
of falling to zero needed time.
Example: Input command “SAFE: STEP 3: IR: TIME: FALL?”
Return message “3.000000E-01”
Description: Return message “3.000000E-01” means setting
voltage value of the main unit STEP 3 fall to zero
needed time is 0.3 sec.
: RANGe
[:UPPer] <numeric_value>
: SOURce: SAFEty: STEP: IR: RANGe: UPPer
This command is in accordance with users’ input current value to
select the range which upper than the current can be measured.
Example: Input command “SAFE: STEP 3: IR: RANG 0.0003”
Description: This command means IR measured current value of the
main unit STEP 3 is set to 300uA. Thus, meanwhile
the selected IR range upper than the current can be
measured is 3mA.
[:UPPer]?
5-28