Remote Operation
4-53
Return Parameters:<aard>, [Unit = Voltage]
Query Example: FETC:ALLV:MAX?
Return Example: 0, 0, 0, 0, 0, 0, 5.12, 0, 12, 0
FETCh:ALLVoltage:MIN?
Type: Channel-Independent
Description: Returns the minimum voltage measured at the input of all load
modules. The return value is 0 when the channel does not exist.
Query Syntax: FETCh:ALLVoltage:MIN?
Return Parameters:<aard>, [Unit = Voltage]
Query Example: FETC:ALLV:MIN?
Return Example: 0, 0, 0, 0, 0, 0, 5.12, 0, 12,0
MEASURE Subsystem 4.3.2.12
MEASure:CURRent?
Type: Channel-Specific
Description: Returns the real time current measured at the load module input.
Query Syntax: MEASure:CURRent?
Return Parameters:<NR2>, [Unit = Ampere]
Query Example: MEAS:CURR?
Return Example: 3.15
MEASure:INPut
Type: Channel-Specific
Description: Selects the input port of the electronic load to measure voltage.
Setting Syntax: MEASure:INPut<space><NRf>
Setting Parameters:<NRf>, LOAD(0), UUT(1)
Setting Example: MEAS:INP LOAD
MEAS:INP 1
Query Syntax: MEASure:INPut?
Return Parameters: <CRD>, LOAD(0), UUT(1)
Query Example: MEAS:INP?
MEASure:POWer?
Type: Channel-Specific
Description: Returns the real time power measured at the load module input.
Query Syntax: MEASure:POWer?
Return Parameters:<NR2>, [Unit = Watt]
Query Example: MEAS:POW?
Return Example: 3.15
MEASure:VOLTage?
Type: Channel-Specific
Description: Returns the real time voltage measured at load module input.
Query Syntax: MEASure:VOLTage?
Return Parameters:<NR2>, [Unit = Volt]
Query Example: MEAS:VOLT?
Return Example: 8.12