EasyManuals Logo

Chroma 63200A Series User Manual

Chroma 63200A Series
141 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #86 background imageLoading...
Page #86 background image
Remote Operation
4-23
Setting Syntax: RESistance:DYNamic:FALL<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: RES:DYN:FALL 2.5 Set falling slew rate to 2.5A/μs.
RES:DYN:FALL 1A/μs Set falling slew rate to 1A/μs.
RES:DYN:FALL MAX Set falling slew rate to the maximum
value of dynamic load.
RES:DYN:FALL MIN Set falling slew rate to the minimum
value of dynamic load.
Query Syntax: RESistance:DYNamic:FALL?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = A/μs]
Query Example: RES:DYN:FALL?
RES:DYN:FALL? MAX
RES:DYN:FALL? MIN
RESistance:DYNamic:IRNG
Type: Channel-Specific
Description: Set the current measurement range in constant resistance dynamic
mode.
Setting Syntax: RESistance:DYNamic:IRNG<space><CRD | NR1>
Setting Parameters:<CRD | NR1>, LOW | L | 0, MIDDLE | M | 1, HIGH | H | 2
Setting Example: RES:DYN:IRNG HIGH Set current range to High.
RES:DYN:IRNG M Set current range to Middle.
RES:DYN:IRNG 0 Set current range to Low.
Query Syntax: RESistance:DYNamic:IRNG?
Return Parameters:<CRD>, LOW(0), MIDDLE(1), HIGH(2) [Unit = None]
Query Example: RES:DYN:IRNG?
VOLTAGE Subsystem 4.3.2.6
VOLTage:STATic:L1
Type: Channel-Specific
Description: Set the static load voltage in constant voltage mode.
Setting Syntax: VOLTage:STATic:L1<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: VOLT:STAT:L1 8 Set voltage of load L1 as 8V.
VOLT:STAT:L1 24V Set voltage of load L1 as 24V.
VOLT:STAT:L1 MAX Set voltage of load L1 as the
maximum value.
VOLT:STAT:L1 MIN Set voltage of load L1 as the minimum
value.
Query Syntax: VOLTage:STATic:L1?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Volt]
Query Example: VOLT:STAT:L1?
VOLT:STAT:L1? MAX
VOLT:STAT:L1? MIN
VOLTage:STATic:L2
Type: Channel-Specific
Description: Set the static load voltage in constant voltage mode.
Setting Syntax: VOLTage:STATic:L2<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: VOLT:STAT:L2 8 Set voltage of load L2 as 8V.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Chroma 63200A Series and is the answer not in the manual?

Chroma 63200A Series Specifications

General IconGeneral
BrandChroma
Model63200A Series
CategoryTest Equipment
LanguageEnglish

Related product manuals