Remote Operation
4-23
Setting Syntax: RESistance:DYNamic:FALL<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: RES:DYN:FALL 2.5 Set falling slew rate to 2.5A/μs.
RES:DYN:FALL 1A/μs Set falling slew rate to 1A/μs.
RES:DYN:FALL MAX Set falling slew rate to the maximum
value of dynamic load.
RES:DYN:FALL MIN Set falling slew rate to the minimum
value of dynamic load.
Query Syntax: RESistance:DYNamic:FALL?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = A/μs]
Query Example: RES:DYN:FALL?
RES:DYN:FALL? MAX
RES:DYN:FALL? MIN
RESistance:DYNamic:IRNG
Type: Channel-Specific
Description: Set the current measurement range in constant resistance dynamic
mode.
Setting Syntax: RESistance:DYNamic:IRNG<space><CRD | NR1>
Setting Parameters:<CRD | NR1>, LOW | L | 0, MIDDLE | M | 1, HIGH | H | 2
Setting Example: RES:DYN:IRNG HIGH Set current range to High.
RES:DYN:IRNG M Set current range to Middle.
RES:DYN:IRNG 0 Set current range to Low.
Query Syntax: RESistance:DYNamic:IRNG?
Return Parameters:<CRD>, LOW(0), MIDDLE(1), HIGH(2) [Unit = None]
Query Example: RES:DYN:IRNG?
VOLTAGE Subsystem 4.3.2.6
VOLTage:STATic:L1
Type: Channel-Specific
Description: Set the static load voltage in constant voltage mode.
Setting Syntax: VOLTage:STATic:L1<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: VOLT:STAT:L1 8 Set voltage of load L1 as 8V.
VOLT:STAT:L1 24V Set voltage of load L1 as 24V.
VOLT:STAT:L1 MAX Set voltage of load L1 as the
maximum value.
VOLT:STAT:L1 MIN Set voltage of load L1 as the minimum
value.
Query Syntax: VOLTage:STATic:L1?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Volt]
Query Example: VOLT:STAT:L1?
VOLT:STAT:L1? MAX
VOLT:STAT:L1? MIN
VOLTage:STATic:L2
Type: Channel-Specific
Description: Set the static load voltage in constant voltage mode.
Setting Syntax: VOLTage:STATic:L2<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: VOLT:STAT:L2 8 Set voltage of load L2 as 8V.