High Power Electronic Load 63200A Series Operation & Programming Manual
4-24
VOLT:STAT:L2 24V Set voltage of load L2 as 24V.
VOLT:STAT:L2 MAX Set voltage of load L2 as the
maximum value.
VOLT:STAT:L2 MIN Set voltage of load L2 as the minimum
value.
Query Syntax: VOLTage:STATic:L2?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Volt]
Query Example: VOLT:STAT:L2?
VOLT:STAT:L2? MAX
VOLT:STAT:L2? MIN
VOLTage:STAT:ILIMit
Type: Channel-Specific
Description: Set the current limit for constant voltage mode.
Setting Syntax: VOLTage:STATic:ILIMit<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example:
VOLT:STAT:ILIM 3 Set the current limit to 3A in
constant voltage mode.
VOLT:STAT:ILIM MAX Set the current limit to the
maximum value in constant
voltage mode.
VOLT:STAT:ILIM MIN Set the current limit to the
minimum value in constant
voltage mode.
Query Syntax: VOLTage:STATic:ILIMit?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Ampere]
Query Example: VOLT:STAT:ILIM?
VOLT:STAT:ILIM? MAX
VOLT:STAT:ILIM? MIN
VOLTage:STATic:RESponse
Type: Channel-Specific
Description: Set the response speed in constant voltage mode.
Setting Syntax: VOLTage:STATic:RESponose<space><NRf>
Setting Parameters:<NRf>, SLOW(0), NORMAL(1), FAST(2)
Example: VOLT:STAT:RES FAST
VOLT:STAT:RES SLOW
Query Syntax: VOLTage:STATic:RESponse?
Return Parameters:<CRD>, SLOW(0), NORMAL(1), FAST(2)
Query Example: VOLT:STAT:RES?
VOLTage:STATic:IRNG
Type: Channel-Specific
Description: Set the current measurement range in constant voltage mode.
Setting Syntax: VOLTage:STATic:IRNG<space><CRD | NR1>
Setting Parameters:<CRD | NR1>, LOW | L | 0, MIDDLE | M | 1, HIGH | H | 2
Setting Example: VOLT:STAT:IRNG HIGH Set current range to High.
VOLT:STAT:IRNG M Set current range to Middle.
VOLT:STAT:IRNG 0 Set current range to Low.
Query Syntax: VOLTage: STATic: IRNG?
Return Parameters:<CRD>, LOW(0), MIDDLE(1), HIGH(2) [Unit = None]
Query Example: VOLT:STAT:IRNG?

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Chroma 63200A Series and is the answer not in the manual?

Chroma 63200A Series Specifications

General IconGeneral
BrandChroma
Model63200A Series
CategoryTest Equipment
LanguageEnglish

Related product manuals