High Power Electronic Load 63200A Series Operation & Programming Manual
4-26
POWer:STATic:FALL
Type: Channel-Specific
Description: Set the falling slew rate of current for constant power mode.
Setting Syntax: POWer:STATic:FALL<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: POW:STAT:FALL 2.5 Set falling slew rate to 2.5A/μs.
POW:STAT:FALL 1A/μs Set falling slew rate to 1A/μs.
POW:STAT:FALL MAX Set falling slew rate to the
maximum value.
POW:STAT:FALL MIN Set falling slew rate to the
minimum value.
Query Syntax: POWer:STATic:FALL?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = A/μs]
Query Example: POW:STAT:FALL?
POW:STAT:FALL? MAX
POW:STAT:FALL? MIN
POWer:STATic:VRNG
Type: Channel-Specific
Description: Set the voltage measurement range in constant power mode.
Setting Syntax: POWer:STATic:VRNG<space><CRD | NR1>
Setting Parameters:<CRD | NR1>, LOW | L | 0, MIDDLE | M | 1, HIGH | H | 2
Setting Example: POW:STAT:VRNG HIGH Set voltage range to High.
POW:STAT:VRNG M Set voltage range to Middle.
POW:STAT:VRNG 0 Set voltage range to Low.
Query Syntax: POWer: STATic: VRNG?
Return Parameters:<CRD>, LOW(0), MIDDLE(1), HIGH(2) [Unit = None]
Query Example: POW:STAT:VRNG?
ADVANCE Subsystem 4.3.2.8
[ADVance:]BATTery:MODE
Type: Channel-Specific
Description: Set run mode in battery discharge mode.
Setting Syntax: [ADVance:]BATTery:MODE<space><CRD | NR1>
Setting Parameters:<CRD | NR1>, CC(0), CR(1), CP(2)
Setting Example: BATT:MODE CC Set run mode = CC
BATT:MODE 2 Set run mode = CP
Query Syntax: [ADVance:]BATTery:MODE?
Return Parameters:<CRD>, CC(0), CR(1), CP(2) [Unit = None]
Query Example: BATT:MODE?
[ADVance:]BATTery:VALue
Type: Channel-Specific
Description: Set load value according to the run mode in battery discharge mode.
Setting Syntax: [ADVance:]BATTery:VALue<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example:
When BATT:MODE set to CC mode, then
BATT:VAL 0.5 Set current = 0.5A
BATT:VAL 500mA Set current = 0.5A
BATT:VAL MAX Set current = maximum value.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Chroma 63200A Series and is the answer not in the manual?

Chroma 63200A Series Specifications

General IconGeneral
BrandChroma
Model63200A Series
CategoryTest Equipment
LanguageEnglish

Related product manuals