Remote Operation
4-15
Return Parameters: <CRD>, OFF, ON [Unit = None]
Query Example: DIO:OUT2?
CURRENT Subsystem 4.3.2.4
CURRent:STATic:L1
Type: Channel-Specific
Description: Set the static load current for constant current static mode.
Setting Syntax: CURRent:STATic:L1<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: CURR:STAT:L1 20 Set the static load parameter L1 =
20A.
CURR:STAT:L1 10A Set the static load parameter L1 =
10A.
CURR:STAT:L1 MAX Set the static load parameter L1 =
maximum value.
CURR:STAT:L1 MIN Set the static load parameter L1 =
minimum value.
Query Syntax: CURRent:STATic:L1?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Ampere]
Query Example: CURR:STAT:L1?
CURR:STAT:L1? MAX
CURR:STAT:L1? MIN
CURRent:STATic:L2
Type: Channel-Specific
Description: Set the static load current for constant current static mode.
Setting Syntax: CURRent:STATic:L2<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: CURR:STAT:L2 20 Set the static load parameter L2 =
20A.
CURR:STAT:L2 10A Set the static load parameter L2 =
10A.
CURR:STAT:L2 MAX Set the static load parameter L2 =
maximum value.
CURR:STAT:L2 MIN Set the static load parameter L2 =
minimum value.
Query Syntax: CURRent:STATic:L2?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Ampere]
Query Example: CURR:STAT:L2?
CURR:STAT:L2? MAX
CURR:STAT:L2? MIN
CURRent:STATic:RISE
Type: Channel-Specific
Description: Set the rising slew rate of current for constant current static mode.
Setting Syntax: CURRent:STATic:RISE<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: CURR:STAT:RISE 2.5 Set rising slew rate to 2.5A/μs.
CURR:STAT:RISE 1A/μs Set rising slew rate to 1A/μs.
CURR:STAT:RISE MAX Set rising slew rate to the
maximum value of static load.