High Power Electronic Load 63200A Series Operation & Programming Manual
Specific Commands 4.3.2
The 63200A series products are equipped with the following specific GPIB commands.
MODE Subsystem 4.3.2.1
MODE
Type: Channel-Specific
Description: This command sets the operational mode for the electronic load.
Setting Syntax: MODE<space><NRf>
Setting Parameters:<CRD>, CCL, CCM, CCH, CRL, CRM, CRH, CVL, CVM, CVH, CPL,
CPM, CPH, CCDL, CCDM, CCDH, CRDL, CRDM, CRDH, BATL,
BATM, BATH, SWDL, SWDM, SWDH, OCPL, OCPM, OCPH,
CCSL, CCSM, CCSH, CZL, CZM, CZH, CVCC, CRCC, CVCR,
AUTO, PROG, MPPTL, MPPTM, MPPTL, MPPTH, UDWL, UDWM,
UDWH, EXTL, EXTM, EXTH
Example: MODE CCL Set CC mode of low range.
MODE CCH Set CC mode of high range.
MODE CCDL Set CC dynamic mode of low range.
MODE CCDH Set CC dynamic mode of high range.
MODE CRL Set CR mode of low range.
MODE CRH Set CR mode of high range.
Query Syntax: MODE?
Return Parameters:<CRD>, CCL, CCM, CCH, CRL, CRM, CRH, CVL, CVM, CVH, CPL,
CPM, CPH, CCDL, CCDM, CCDH, CRDL, CRDM, CRDH, BATL,
BATM, BATH, SWDL, SWDM, SWDH, OCPL, OCPM, OCPH,
CCSL, CCSM, CCSH, CZL, CZM, CZH, CVCC, CRCC, CVCR,
AUTO, PROG, MPPTL, MPPTM, MPPTL, MPPTH, UDWL, UDWM,
UDWH, EXTL, EXTM, EXTH
Query Example: MODE?
LOAD Subsystem 4.3.2.2
LOAD[:STATe]
Type: Channel-Specific
Description: The LOAD command makes the electronic load active/on or
inactive/off.
Setting Syntax: LOAD[:STATe]<space><NRf>
Setting Parameters:<NRf>, OFF(0), ON(1)
Setting Example: LOAD ON Activate the electronic load.
LOAD 0 Inactivate the electronic load.
Query Syntax: LOAD[:STATe]?
Return Parameters:<CRD>, OFF, ON
Query Example: LOAD?
LOAD:PROTection?
Type: Channel-Specific
Description: This command returns the status of electronic load.
Setting Syntax: None
Setting Parameters:None