Remote Operation
BATT:VAL MIN Set current = minimum value.
When BATT:MODE set to CR mode, then
BATT:VAL 0.5 Set resistance = 0.5Ω.
BATT:VAL 500mΩ Set resistance = 0.5Ω.
BATT:VAL MAX Set resistance = maximum value.
BATT:VAL MIN Set resistance = minimum value.
When BATT:MODE set to CP mode, then
BATT:VAL 0.5 Set power = 0.5W.
BATT:VAL 500mW Set power = 0.5W.
BATT:VAL MAX Set power = maximum value.
BATT:VAL MIN Set power = minimum value.
Query Syntax: [ADVance:]BATTery:VALue?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Ampere | Ohm | Watt]
Query Example: BATT:VAL?
BATT:VAL? MAX
BATT:VAL? MIN
[ADVance:]BATTery:RISE
Type: Channel-Specific
Description: Set rising slew rate of current in battery discharge mode.
Setting Syntax: [ADVance:]BATTery:RISE<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: BATT:RISE 0.1 Set slew rate = 0.1A/μs
BATT:RISE 100mA/μs Set slew rate = 0.1A/μs
BATT:RISE MAX Set slew rate = maximum value.
BATT:RISE MIN Set slew rate = minimum value.
Query Syntax: [ADVance:]BATTery:RISE?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = A/uS]
Query Example: BATT:RISE?
BATT:RISE? MAX
BATT:RISE? MIN
[ADVance:]BATTery:FALL
Type: Channel-Specific
Description: Set falling slew rate of current in battery discharge mode.
Setting Syntax: [ADVance:]BATTery:FALL<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: BATT:FALL 0.1 Set slew rate = 0.1A/μs
BATT:FALL 100mA/μs Set slew rate = 0.1A/μs
BATT:FALL MAX Set slew rate = maximum value.
BATT:FALL MIN Set slew rate = minimum value.
Query Syntax: [ADVance:]BATTery:FALL?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = A/uS]
Query Example: BATT:FALL?
BATT:FALL? MAX
BATT:FALL? MIN
[ADVance:]BATTery:ENDVoltage
Type: Channel-Specific
Description: Set end voltage for battery discharge mode.
Setting Syntax: [ADVance:]BATTery:ENDVoltage<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.