High Power Electronic Load 63200A Series Operation & Programming Manual
4-22
RESistance:DYNamic:T2
Type: Channel-Specific
Description: Set duration parameter T2 for constant resistance dynamic mode.
Setting Syntax: RESistance:DYNamic:T2<space><NRf+>[suffix]
Setting Parameters:<NRf+>, 10μs ~ 100s, Resolution = 10μs, Unit = Second
Setting Example: RES:DYN:T2 10ms Set the dynamic duration T2 = 10ms.
RES:DYN:T2 2 Set the dynamic duration T2 = 2s.
RES:DYN:T2 MAX Set the dynamic duration T2 as
maximum value.
RES:DYN:T2 MIN Set the dynamic duration T2 as
minimum value.
Query Syntax: RESistance:DYNamic:T2?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Second]
Query Example: RES:DYN:T2?
RES:DYN:T2? MAX
RES:DYN:T2? MIN
RESistance:DYNamic:REPeat
Type: Channel-Specific
Description: Set the repeat count for constant resistance dynamic mode.
Setting Syntax: RESistance:DYNamic:REPeat<space><NRf+>
Setting Parameters:<NRf+>, 0 ~ 65535, Resolution = 1, Unit = None
Setting Example: RES:DYN:REP 500 Set repeat count = 500
RES:DYN:REP MAX Set repeat count = maximum value.
RES:DYN:REP MIN Set repeat count = minimum value.
Query Syntax: RESistance:DYNamic:REPeat?[<space><MAX | MIN>]
Return Parameters:<NR1>, [Unit = None]
Query Example: RES:DYN:REP?
RES:DYN:REP? MAX
RES:DYN:REP? MIN
RESistance:DYNamic:RISE
Type: Channel-Specific
Description: Set the rising slew rate of current for constant resistance dynamic
mode.
Setting Syntax: RESistance:DYNamic:RISE<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: RES:DYN:RISE 2.5 Set rising slew rate to 2.5A/μs.
RES:DYN:RISE 1A/μs Set rising slew rate to 1A/μs.
RES:DYN:RISE MAX Set rising slew rate to the maximum
value of dynamic load.
RES:DYN:RISE MIN Set rising slew rate to the minimum
value of dynamic load.
Query Syntax: RESistance:DYNamic:RISE?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = A/μs]
Query Example: RES:DYN:RISE?
RES:DYN:RISE? MAX
RES:DYN:RISE? MIN
RESistance:DYNamic:FALL
Type: Channel-Specific
Description: Set the falling slew rate of current for constant resistance dynamic
mode.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Chroma 63200A Series and is the answer not in the manual?

Chroma 63200A Series Specifications

General IconGeneral
BrandChroma
Model63200A Series
CategoryTest Equipment
LanguageEnglish

Related product manuals