High Power Electronic Load 63200A Series Operation & Programming Manual
4-56
Setting Syntax: PROGram:DATA:STEP<space><Arg1>,<Arg2>,<Arg3>,<Arg4>,
<Arg5>,<Arg6>,<Arg7>,<Arg8>,<Arg9>,<Arg10>,<Arg11>,
<Arg12>,<Arg13>,<Arg14>,<Arg15>,<Arg16>
Setting Parameters:
Selects a program to be set:
Arg1: <NR1>, 1 ~ 10, Resolution = 1, Unit = None.
Set the trigger mode of sequence:
Arg2: <NRf>, SKIP(0), AUTO(1), MANUAL(2), EXTERNAL(3),
Unit = None.
Set the run mode of sequence:
Arg3: <NRf>, CC(0), CR(1), CV(2), CP(3), Unit = None.
Set the mode’s range of sequence:
Arg4: <NRf>, LOW(0), MIDDLE(1), HIGH(2), Unit = None.
Set the start value according to run mode in sequence:
Arg5: <NRf>, Refer to respective specification for valid value
range.
Set the end value according to run mode in sequence:
Arg6: <NRf>, Refer to respective specification for valid value
range.
Set the falling of slew rate in sequence:
Arg7: <NRf>, Refer to respective specification for valid value
range.
Set the rising of slew rate in sequence:
Arg8: <NRf>, Refer to respective specification for valid value
range.
Set the dwell time of sequence:
Arg9: <NRf>, 0.1ms ~ 30s, Resolution = 0.0001s, Unit =
Second.
Set the high-level of voltage specific in sequence:
Arg10: <NRf>, Refer to respective specification for valid value
range.
Set the low-level of voltage specific in sequence:
Arg11: <NRf>, Refer to respective specification for valid value
range.
Set the high-level of current specific in sequence:
Arg12: <NRf>, Refer to respective specification for valid value
range.
Set the low-level of current specific in sequence:
Arg13: <NRf>, Refer to respective specification for valid value
range.
Set the high-level of power specific in sequence:
Arg14: <NRf>, Refer to respective specification for valid value
range.
Set the low-level of power specific in sequence:
Arg15: <NRf>, Refer to respective specification for valid value
range.
Set the delay time of Pass/Fail in sequence:
Arg16: <NRf>, 0s ~ 30s, Resolution = 0.0001s, Unit = Second.
Setting Example: PROG:DATA:STEP ,AUTO,CC,2,3.5,20.0,0.5,0.5,2,-1,-1,-1,-1,-1,-
1,1
Query Syntax 1: PROGram:DATA:STEP?<space><Arg1>[<space><MAX | MIN>]
Selects a program:
Arg1: <NR1>, 1 ~ 10, Resolution = 1, Unit = None.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Chroma 63200A Series and is the answer not in the manual?

Chroma 63200A Series Specifications

General IconGeneral
BrandChroma
Model63200A Series
CategoryTest Equipment
LanguageEnglish

Related product manuals