High Power Electronic Load 63200A Series Operation & Programming Manual
PROGram:SEQuence:CLEar
Type: Channel-Specific
Description: Clear all sequence in program file what specified.
Setting Syntax: PROGram:SEQuence:CLEar<space><NR1>
Setting Parameters:<NR1>, 1 ~ 10, Resolution = 1, Unit = None
Setting Example: PROG:SEQ:CLE 3
Query Syntax: None
Return Parameters:None
Query Example: None
PROGram:SEQuence:FAIL?
Type: Channel-Specific
Description: This command returns the fail of sequence in specification.
Setting Syntax: None
Setting Parameters:None
Query Syntax: PROGram:SEQuency:FAIL?
Return Parameters:<aard>, xx-xxx,xx-xxx,xx-xxx…etc, which front of “-“ is the program
number and rear of “-“ is the sequence number.
Query Example: PROG:SEQ:FAIL?
Return Example: 1-2,5-13,10-8
PROGram:SEQuence:REMain
Type: Channel-Specific
Description: This command returns the remains sequence of unused.
Setting Syntax: None
Setting Parameters:None
Query Syntax: PROGram:SEQuency:REMain?
Return Parameters:<NR1>
Query Example: PROG:SEQ:REM?
PROGram:SEQuence:SPECification?
Type: Channel-Specific
Description: This query returns the specification of program sequence from the
DC Electronic Load in binary format. The specification either
voltage、current or power are consist of total sequences in program
file that user specified in format of 32bits float point.
Setting Syntax: None
Setting Parameters:None
Query Syntax: PROGram:SEQuency:SPECification?<space><Arg1>,<Arg2>
Selects a program:
Arg1: <NR1>, 1 ~ 10, Resolution = 1, Unit = None.
Selects a parameter of measurement:
Arg2: <NRf>, V(0), I(1), P(2), Unit = None.
Return Parameters: <DLABRD>, for example: when the 12 sequences in program 1,
than the return will be the format show in below.