High Power Electronic Load 63200A Series Operation & Programming Manual
4-20
RES:STAT:L2? MIN
RESistance:STATic:RISE
Type: Channel-Specific
Description: Set the rising slew rate of current for constant resistance mode.
Setting Syntax: RESistance:STATic:RISE<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: RES:STAT:RISE 2.5 Set rising slew rate to 2.5A/μs.
RES:STAT:RISE 1A/μs Set rising slew rate to 1A/μs.
RES:STAT:RISE MAX Set rising slew rate to the
maximum value of static load.
RES:STAT:RISE MIN Set rising slew rate to the
minimum value of static load.
Query Syntax: RESistance:STATic:RISE?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = A/μs]
Query Example: RES:STAT:RISE?
RES:STAT:RISE? MAX
RES:STAT:RISE? MIN
RESistance:STATic:FALL
Type: Channel-Specific
Description: Set the falling slew rate of current for constant resistance mode.
Setting Syntax: RESistance:STATic:FALL<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: RES:STAT:FALL 2.5 Set falling slew rate to 2.5A/μs.
RES:STAT:FALL 1A/μs Set falling slew rate to 1A/μs.
RES:STAT:FALL MAX Set falling slew rate to the
maximum value of static load.
RES:STAT:FALL MIN Set falling slew rate to the
minimum value of static load.
Query Syntax: RESistance:STATic:FALL?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = A/μs]
Query Example: RES:STAT:FALL?
RES:STAT:FALL? MAX
RES:STAT:FALL? MIN
RESistance:STATic:IRNG
Type: Channel-Specific
Description: Set the current measurement range in CR mode.
Setting Syntax: RESistance:STATic:IRNG<space><CRD | NR1>
Setting Parameters:<CRD | NR1>, LOW | L | 0, MIDDLE | M | 1, HIGH | H | 2
Setting Example: RES:STAT:IRNG HIGH Set current range to High.
RES:STAT:IRNG M Set current range to Middle.
RES:STAT:IRNG 0 Set current range to Low.
Query Syntax: RESistance:STATic:IRNG?
Return Parameters:<CRD>, LOW(0), MIDDLE(1), HIGH(2) [Unit = None]
Query Example: RES:STAT:IRNG?
RESistance:DYNamic:L1
Type: Channel-Specific
Description: Set the load resistance during T1 period for constant resistance
dynamic mode.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Chroma 63200A Series and is the answer not in the manual?

Chroma 63200A Series Specifications

General IconGeneral
BrandChroma
Model63200A Series
CategoryTest Equipment
LanguageEnglish

Related product manuals