FSG/MTM/001 060511
87
Fig. 4-50 Focus Pattern, 25”, page 127
Fig. 4-51 Focus Pattern, 40”, page 128
Fig. 4-52 Focus Pattern, 54”, page 129
In the description of the adjustment procedures, references are made to sub-patterns A, B, and
C of the Optical Adjustment Pattern. These sub-patterns are shown in Fig. 4-45, page 122.
4.2.1 Check of Light Profiles
Images of the Light Profiles are shown in:
Fig. 4-5 COUGAR 25, CHAMELEON 25, Light Profiles Uncorrected , page 89
Fig. 4-6 COUGAR 25, CHAMELEON 25, Light Profiles Corrected , page 89
Fig. 4-7 COUGAR 36, CHAMELEON 36, PREMIER 36, Light Profiles Uncorrected, page
90
Fig. 4-8 COUGAR 36, CHAMELEON 36, PREMIER 36, Light Profiles Corrected, page 90
Fig. 4-9 CHROMA Tx 40, CRYSTAL Tx 40, Light Profiles Uncorrected, page 91
Fig. 4-10 CHROMA Tx 40, CRYSTAL Tx 40, Light Profiles Corrected, page 91
Fig. 4-11 MAGNUM XL 54, Light Profiles Uncorrected, page on page 92
Fig. 4-12 MAGNUM XL 54, Light Profiles Corrected, page 92
Fig. 4-13 COUGAR Tx 36, CHAMELEON Tx 36, PREMIER Tx 36, Light Profiles
Uncorrected, page 93
Fig. 4-14 COUGAR Tx 36, CHAMELEON Tx 36, PREMIER Tx 36, Light Profiles
Corrected, page 93
Procedure for check of Light Profiles
Place the Original Guide Plate in the scan-area.
Start SCANtest 6
Select Test 9
Select Uncorrected or Corrected Light Profile
Press Execute
Select Green (Default for Uncorrected)
Select Detail View: Light Profile
Select Camera
Compare the image of the Software Oscilloscope with the relevant sample image listed above.
If the signal levels of the uncorrected Light Profiles differ much from the sample images, the
scanner might need to be calibrated by Scanner Maintenance.