Chapter V Troubleshooting
5-1
C
996-0255 rev. b
No lasing output
Lamps
flash?
Go to free run &
measure IR energy
see page 5-10
Perform electronic cabinet
troubleshooting procedure
see page 5-6
yes no
yes no
Perform Q-switch
electronics test
see page 5-8
IR power
normal?
1) Check for optical damage
2) Adjust rear mirror
3) Change flashlamp
see page 4-2 & 3
Troubleshooting guide for
Precision series