Precision II Operation and Maintenance Manual
C
2-2
996-0255 rev. b
10) Select desired programmed mode of operation by press-
ing the PROGRAM UP key on the RB601.
11) Once the PGM is located press ACTIVATE key to engage
the desired PGM.
12) For lasers with the SI-500 option (Seeded SLM), flip the
shutter switch to the OPEN position on the seed laser's
control panel .
13) On the RB601 press on the keys SHUTTER and Q-
SWITCH. The four LEDs across the top of the RB601
should be on or flashing.
14) Open desired wavelength port and maximize harmonic
crystals if necessary.
B. Shutdown procedure
1) Press the STOP key on the RB601.
2) Close all wavelength ports.
3) For lasers with the SI-500 option (seeded SLM), turn the
key switch to the STANDBY position and the shutter
switch to the CLOSED position on the seed laser's control
panel.
4) Turn key on front panel of CU601C clockwise to the OFF
position.
Caution:
Leaving the seeder in the ON position reduces Seed
Laser life time and will result in a premature failure of
the pumping diode.
5) Leave the 24V power supply switch on rear of electronics
rack in the ON position.
6) Switch off electronics rack circuit breaker located on front,
upper right hand side of cabinet.
7) Close city water and drain source connected to laser
system.