EasyManua.ls Logo

Continuum Precision II Series - Page 260

Continuum Precision II Series
274 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
38
POSSIBLE CAUSE:
External cooling water has frequent and large pressure fluctuations.
REMEDY:
Check external cooling water for pressure fluctuations. Install pressure regulator set at 20 psi
if necessary.
SYMPTOM #6: Q-switch BUT Reduction Changes (slowly) by more than 10% after external
cooling water is flowing.
POSSIBLE CAUSE:
Etalon effects from various coated and uncoated surfaces within the resonator.
REMEDY:
With the Mode Switch in MANUAL, change the Laser Frequency Potentiometer (located on
top of the laser head) while observing the reduction in Q-switch BUT. As the seed laser is
tuned in and out of resonance with the host, if the Q-switch BUT periodically changes from
a large BUT to a smaller BUT, some element within the host cavity is acting as an etalon.
This should be tracked down and corrected by tilting the element off axis. This would
happen every 5 to 25 resonances of the seeder with the host, and is not to be confused with
the mode hopping of the seed laser which happens after more than 100 resonances or >1cm
-1
of seed laser tuning.
SYMPTOM #7: No output from the seed laser.
POSSIBLE CAUSE:
Laser diode has reached the end of its life or the seeder NPRO has misaligned.
REMEDY:
The unit must be returned to Lightwave for repair.

Table of Contents