The CU601C will beep two long beeps during a start up when the battery
has failed or been replaced. After a battery failure or a new battery has
been installed, the CU601C will default the active parameters to the
following values:
All 16 user programs will default to the following values:
The rep rate max. and rep rate min. limits will default to the following
values:
The rep rate max. and rep rate min. values will have to be reset. Call
Continuum for help. The RS232 baud rate will default to:
The host RS232 baud rate may have to be set to a new value if the host
computer program is different from the default baud rate.
The last PU will default to 1.
REP RATE 10Hz
Q-SWITCH DELAY 1uS
PULSE DIVISION F/1
SHOT COUNT 0
REP RATE 10Hz
Q-SWITCH DELAY 1uS
PULSE DIVISION F/1
REP RATE MIN 1Hz
REP RATE MAX 15Hz
BAUD RATE 9600