4. During this phase, before the reference cell is subjected to any load, zero the reading
by pressing the key;
NOTE:
Do not zero the load reading once the load cell has been subjected to loading.
NOTE:
The key can be pressed several times to zero the load reading on the display
before the frame start the loading phase. The key will not be active if an
increase of at least 0.25% of the full scale is detected by the load sensor.
5. When the load cell is close to the upper platen of the compression machine it is
recommended that the load rate is decreased so as to not apply a shock load to the
cell. To do this set lever A up, turn lever B slowly forward to control the speed of the
piston with the flow control valve;