Manuale di Istruzione
Instruction Manual
23/08/2018 Rev.1 EN
74
4.3.3 Testing a sample
1. Switch ON the machine and wait 15 minutes warm up time;
2. Place the sample in the test frame and close front door and rear guards (see previous
WARNIGS);
3. After accessing TEST RUN menu, if two channels have been activated into the
system (see chapter 6.1.1 for channel activation), use the keys to
select the desired channel and press the key to confirm it.
Fig. 4-21
Otherwise the system will directly go to the following screen;
Fig. 4-22
4. Set the test parameters as needed and press the key to proceed, the system
will require confirmation to save the changes:
Fig. 4-23