vPad-A1 OPERATORS MANUAL
vPad-A1 Simulator
Datrend Systems Inc.
File: P24601_20160926_125522_P.txt
Date: 2016 Sep 26
Time: 12:55:22
vPad ID:
vPad-A1: VBU-DEMO-BU,2016-09-26
VSM-DEMO-PS,2016-09-26
VSP-DEMO-O2,2016-09-26
VBP-DEMO-BP,2016-09-26
Equipment Information...
Control Number: P24601
Description: PATIENT MONITOR
Manufacturer: MED CORP
Model: S9001
Serial Number: 30624700
Location: CARDIAC CARE UNIT
Facility: CITY GENERAL
Technician ID: RW
Tech Time (hrs): 0.6
(min): 35
Overall Result: PASS
Signature:____________________________ Date:__________
Detailed Test Data...
_______________________________________________________
Task: Normal sinus rhythm
Axis Deviation: Intermediate (Normal) Axis
Amplitude: 2.0 mV
Rate: 80 BPM
Artifact: OFF
Mode: Adult
Data: Amplitude: 2.2 mV
Rate: 79 BPM
Amplitude estimated from monitor display.
Result: PASS
_______________________________________________________
See test comment.
END OF RECORD
Appendix A # Page 92