vPad-A1 OPERATORS MANUAL
1.3 Pulse Oximetry (SpO2) Testing
Pulse oximetry or SpO2 testing is provided by the vPad-O2 module of vPad-A1.
Saturation (SpO2):
• Range: 30 to 100% SpO2*
• Increments: 1%
• Accuracy: ±1 count + specified accuracy of the DUT
when within DUT specified range
• Presets: 6, user-definable
* Range of adjustment and presets may vary according to pulse oximeter specifications
Heart Rate:
• Range: 20 to 300 BPM
• Increments: 1BPM
• Accuracy: ±0.25 BPM in sync mode
±1 BPM otherwise
• Presets: 6, user-definable
Pulse Amplitude:
• Range: 0 to 100%
• Increments: 1% steps
• Accuracy: ±1 %
• Presets: 6, user-definable
Specifications/Chapter 1 # Page 9