vPad-A1 OPERATORS MANUAL
It can be set from 0.05 to 1.0 ohms in 0.05 ohm increments, and from 1.0 to 5.0 ohms in 0.25 ohm
increments.
4.3.2.4 Lead Selection
The Leads parameter selects the ECG leads across which the simulated Impedance Variation is
applied. Choices are between the RA and LL leads, or alternatively between the RA and LA leads.
4.3.2.5 Inspiration/Expiration Ratio
The
Inspiration/Expiration Ratio (I:E) defines a relationship between inspiration (inhalation) time and
expiration (exhalation) time, which is an indicator of patient respiratory effort. It can be set to one
of the following options:
1:1, 1:2, 1:3, 1:4, 1:5, or Ventilated 1:1
4.3.2.6 Apnea
Apnea is the temporary cessation of breathing, which may occur while sleeping or under anaesthesia.
This setting imitates the condition by suspending Respiration simulations for the selected duration.
Available settings are:
Off, Continuous, 12 seconds, 22 seconds, or 32 seconds
Note
: ‘Continuous’ means breathing is stopped until another Apnea setting is chosen.
Manual Operation/Chapter 4 # Page 42