vPad-A1 OPERATORS MANUAL
When creating an AutoSetting, you have an opportunity to create an
Instruction that will be visible to the user when on the Results screen. Press
the
Edit Instr. button to make the Instructions dialog visible. Touch inside
the white text box to activate the keyboard and enter the instruction text.
Press
Save to retain the instruction. Pressing Close is equivalent to Cancel
and no changes will be retained.
When the Results button is pressed, the Instructions (if any) will be shown
on the screen. If you wish to enter any comments for the measurement being
recorded, select Comments and touch inside the white text window.
To delete an AutoSetting, press the icon after it is selected, then use the button.
5.2.2 AutoSequences
An
AutoSequence is a pre-defined series of simulation or performance tests. These tests may involve only
one parameter or multiple A1 simulation modules acting in unison.
Each step of an AutoSequence is called a Group, which consists of one or more AutoSettings pulled from
various simulation modules (see 5.2.1 -
AutoSettings
). For instance, a particular AutoSequence Group may
define an
ECG state (Waveform, Heart Rate, Amplitude, etc.) as well as a Dynamic IBP waveform and a
saved NIBP Simulation, all at the same time. The Group applies these definitions concurrently, as if they
have been simultaneously recalled through their respective AutoSettings menus.
Figure 121
Figure 122
Automation and Test Records/Chapter 5 # Page 80