vPad-A1 OPERATORS MANUAL
5 Automation and Test Records
This chapter explains how to use vPad-A1 to perform testing of multi-parameter patient monitors, pulse
oximeters, and/or NIBP monitors, following a predefined sequence of steps referred to as an
AutoSequence. The generation and management of Test Results and Test Records are also covered in this
chapter.
5.1 Automated Performance Testing
An important element in the maintenance of patient-monitoring equipment is the periodic verification of
the ability of a Device Under Test (DUT) to detect, measure, display, and/or interpret patient vital signs. The
simulator or performance tester’s role is to generate precise, calibrated signals that produce predictable
results in a functional monitoring device.
A simulation or performance test would typically follow these four steps:
Setting up the desired simulations (ECG, Respiration, SpO2, etc.) in the A1 App
Connecting the DUT to A1 module(s) using signal leads, cables, hoses and adapters
Observing the DUT for vital signs readings, measurements or displays
Assigning a “PASS” or “FAIL”
Test Result based on standards and tolerances
After generating a series of Test Results for the same DUT, the A1 can consolidate them into an equipment
Test Record that documents in detail the conditions and outcomes of the tests performed. Furthermore,
these Test Records can be integrated with another Equipment Management System, depending on
institutional or regulatory requirements.
Automation and Test Records/Chapter 5 # Page 73