vPad-A1 OPERATORS MANUAL
Pressure Source:
• Range: 10.0 to 400.0 mmHg
• Accuracy: ± 0.5 mmHg
• Resolution : 0.1 mmHg
• Units: mmHg
• Test Modes: Automatic
Manual
Step (definable)
• User-defined Settings: Unlimited
Leak Test:
• Range: 0 to 200mmHg/min
• Inflation: Automatic or Manual
• Pressure: 20 to 400mmHg
• Elapsed Time: 30 to 600 sec
• Presets: Pressure, Elapsed Time
6 each, user-definable values
• User-defined AutoSettings: Unlimited
Over PressureTest:
• Inflation: Automatic or Manual
• Range: 20 to 400mmHg
• Release Time: 1-999 sec
• Presets: Pressure, Release Time
6 each, user-definable values
• User-defined AutoSettings: Unlimited
Specifications/Chapter 1 # Page 12