17
AUTOMATIC CONSTANTS UPDATE
Selecng the [AUTOMATIC CONSTANTS UPDATE] checkbox enables the Automac Constants
Update feature. This feature automacally calculates and updates the personalized lens
constant for the selected doctor and lens based on the records in the table that have “Use
Record” selected. The Personalized Lens Constants are also automacally recalculated when
new records for the specied doctor and IOL manufacturer and model are created and the “Add
to Personalized Lens Table” checkbox on the “Paent Data” Screen is checked. The number of
records used to calculate the constants are also counted and stored automacally upon each
recalculaon. Automacally Calculated Personalized Constants are only available for use once
the number of selected records exceeds the “Min Cases Before Using Personalized Constants”
set in Doctor Preferences.
Min Cases: This value is displayed on the Personalized Constants table. This
indicates the minimum number of cases used in calculang personalized
constants and is not editable on this screen. To edit this value, go to the
Doctor Preferences menu.
Selected Cases: This value is a counter that noes the user how many records have
been selected in the Use Record column. It cannot be edited on this
screen, but will update automacally as records are selected or
deselected for use in personalized constant calculaons.
OPERATOR PREFERENCES MENU
The Operator Preferences Menu allows the user to create, edit or delete individual operators for
the device. Operator preferences such as video playback speeds and auto-saving opons can
also be set in this menu. A eld is provided for user to record notes associated with the
operator. Other sengs that can be adjusted are listed below.
Video Buer Frames: The video buer determines the amount of video that is captured
before recording is stopped or overwrien. Increasing the number
of frames allows the user to record longer videos. This will also
result in larger le sizes.
Bidireconal Scan: If enabled, the B-Scan probe will acquire image informaon
in both sweep direcons.
Average / Interpolate: The soware can be congured to use a frame-to-frame averaging
and/or interpolaon algorithm. Using these algorithms may improve
image quality depending upon the system processor.
.
SYSTEM PREFERENCES MENU
The System Preferences Menu allows the user to set the pracce idencaon informaon, date
format and database login conguraon. Other sengs that can be adjusted are listed below.
Voice Control: Voice control can be enabled/disabled in this menu. Refer To the
B-Scan/UBM Voice Control and A-Scan Voice Control secons for more
informaon.
Default Keratometer The Default Keratometer Index (nc) must be set to match the nc value
Index (nc) the keratometer used. This must be done prior to performing IOL
Calculaons.