POST REFRACTIVE CALCULATIONS
NOTE: The corrected corneal power (Kcorr) calculated using the formulas described are NOT to
be interpreted as a suggested or recommended corneal power on the part of DGH
Technology, Inc. The calculated corrected corneal power (Kcorr) should be used as a
guideline only. It is based upon the accuracy of the post refracve formula used and the
data entered. The determinaon of the corrected corneal power (Kcorr) to use for
performing IOL Power Calculaons must be made by the surgeon, based on personal
experiences and previous postoperave results.
1. Select the “Post Refracve” checkbox on the IOL Calculator Screen next to the eye that has
previously undergone corneal refracve surgery. This will automacally open the Post
Refracve Calculator.
NOTE: The Post Refracve Calculator can be opened again at any me by selecng the
[KCORR] buon on the IOL Calculator Screen. The [KCORR] buon is inacve unl
the “Post Refracve” checkbox has been selected.
2. When the “Post Refracve” checkbox is selected, the Pre-Op group box on the IOL
Calculator screen will change in the following ways:
- K1 and K2 will be replaced with Kcorr or Pre Ref. K and Post Ref. K (Pre Refracve
Surgery and Post Refracve Surgery) depending on the Post Refracve Method selected.
- The values Kcorr, Pre Ref. K, and Post Ref. K are “read only”
- K1 and K2 are entered automacally into the “Post Refracve Surgery” secon of the Post
Refracve Calculator
NOTE: Deselecng the “Post Refracve” checkbox returns all elds to the original
conguraon.
3. In the Post Refracve Calculator, select the desired eye(s) on which to perform Post
Refracve calculaons using the “Calculate OD” and “Calculate OS” checkboxes.
4. Select the desired Post Refracve formula using the method drop-down box. The required
inputs for the selected method will then be highlighted in white.
5. Enter all known inputs. If a eld is gray, that input is not required for the selected method,
but can sll be edited. K1 and K2 values can be entered as either mm or D, but both K
values in a secon must be in the same units. Pre-Surgical and Post-Surgical K values can be
entered in dierent units.
NOTE: The refracve index (nc) displayed in the IOL Calculator will be automacally
transferred to both the Pre-Surgical and Post-Surgical secons of the Post Refracve
Calculator. If a dierent keratometer was used for Pre-Surgical measurements, the
correct nc for the keratometer used must be entered in the Post Refracve
Calculator.
6. Click the [CALCULATE] buon and the corrected corneal power (Kcorr) will be calculated. If
required inputs are missing, the empty eld will be highlighted in red. All elds must be
entered to complete the calculaon. A successful calculaon will display the Kcorr in the
“Outputs” secon of the calculator. Kcorr is given in diopters (D) using a keratometer index
(nc) of 1.3375.