PERFORMING IOL CALCULATIONS
NOTE: The calculated IOL powers are NOT to be interpreted as suggested or recommended
implant powers on the part of DGH Technology, Inc. The calculated IOL should be used as
a guideline only. It is based upon the accuracy of the IOL formula used and the data
entered. The nal IOL implant power selecon must be made by the surgeon, based on
personal experiences and previous post-operave results.
1. Select the “Paent Data” Screen and either enter a new paent or select an exisng
paent.
2. Select the “A-Scan” Screen tab and obtain A-Scan measurements for the paent. Refer to
the Performing A-Scans (Contact) or Performing A-Scans (Immersion) secons for
informaon on using the Scanmate soware for these procedures. Verify that the
informaon is correct and enter any missing Pre-Operave Data.
3. Select the Doctor that is performing the Lens Replacement Surgery using the drop down
box. The preferred IOLs and formulas will automacally be loaded for the selected Doctor.
4. If the Paent has previously undergone corneal refracve surgery, select the “Post
Refracve” checkbox for the aected eye or for both eyes. This will automacally open the
Post Refracve Calculator. Refer to the Post Refracve Calculaons secon for more
informaon.
5. Select the measurement source. Choices include A-Scan average, A-Scan Single
Measurement and User Input. If user input is selected, complete the highlighted elds.
6. The Corneal Refracve Index (nc) of the keratometer used to calculate the corneal power
(K1 and K2) for the current paent should be entered in this eld. This eld will
automacally be completed with the default corneal refracve index (nc) set in System
Preferences, but it can be changed by the user on this screen.
WARNING: Failure to adjust the Default Keratometer Index (nc) to match the actual nc of
the keratometer used to measure the eye may result in errors in the IOL Calculaon .
7. If desired, modify the IOLs and formulas using the appropriate drop-down boxes. The
“default” lens constant for the selected IOL and formula will be shown underneath the
formula selecon box.
8. If desired, select the “Personalized Constant” checkbox. This opon will only be available if
a personalized lens constant has been calculated for the currently selected Doctor and IOL.
When selected, the personalized lens constant will replace the default lens constant. Refer
to the Personalized Lens Constants secon for more informaon.
9. Once all necessary inputs are entered, the DGH Scanmate will perform IOL calculaons
using the selected formula for each of the six (6) selected IOLs on the screen. Any
necessary inputs for performing an IOL Power Calculaon that are either missing or out of
range are automacally highlighted in red.
10. Save the pre-operave data for the current paent and procedure by selecng the [SAVE]
icon.