EasyManua.ls Logo

DGH TECHNOLOGY SCANMATE FLEX - Performing A-Scans (Contact); Contact Measurement Preparation; Contact Scan Execution; Contact Measurement Alignment

Default Icon
64 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
29
PERFORMING A-SCANS (CONTACT)
WARNING: Exercise cauon when coming into contact with a paents eye. To avoid
potenal injury, use minimal pressure when applying the probe or adjusng its posion.
Only make minor posion adjustments while in contact with the eye. If signicant posion
adjustment is necessary, remove the probe from the eye and then reapply.
WARNING: Before each procedure, the ultrasound probe must be cleaned and disinfected.
Refer to the Cleaning and Disinfecon secon.
NOTE: Refer to the Alignment Ranking System and Video Buer Full secons for more
informaon on these topics.
1. Create a new paent or select an exisng paent on the Paent Screen.
2. Select the A-Scan Screen tab. If necessary, click the [CHANGE EYE] buon to change the
eye being measured (OD or OS).
3. Select the Contactradio buon on the right-hand side of the screen. Select the name of
the Operator performing the measurement from the drop-down box.
NOTE: If the operator is not listed in the Operatordrop-down box, add the new operator
by selecng Preferences > Operator and clicking the [NEW] buon.
4. Select the correct Lens and Vitreous type for the eye to be measured.
5. Adjust the compression lockout sensivity track bar (Comp) to the desired seng. Higher
Compression Sensivity will decrease the amount of error due to corneal compression, but
will make it more dicult to obtain measurements.
NOTE: Refer to Page 27 for more informaon on Corneal Compression Sensivity.
6. Seat or recline the paent in a posion comfortable for both Paent and Operator. Head
movement must be minimized during the exam. Anestheze the eye and have the paent
xate with both eyes on a spot on the ceiling or wall.
7. To Start scanning†: - Click the [START SCAN] buon; or
- Press the Space Bar (on the keyboard); or
- Press the Foot Pedal (Single Pedal accessory); or
- Press the Le Foot Pedal (Double Pedal accessory)
A waveform will be displayed on the screen, but the unit will not begin recording.
8. Stabilize the hand holding the probe to minimize indentaon of the cornea or
excessive movement. Typically, this can be done by resng a palm on the cheek or
forehead of the paent.
9. The transducer should be held perpendicular to the surface of the eye as it approaches the
cornea. Observe the approach from a vantage point that allows a view across the
topography of the eye. This method aids in seeing the exact moment when the probe
contacts the cornea, allowing the user to minimize indentaon of the cornea.
10. Gently apply the probe p to the center of the cornea. The unit will start beeping at the
moment of contact, which indicates coupling has been achieved. Minimize indentaon of
the cornea.
NOTE: Stopped or intermient beeping indicates that the probe p has been removed from
contact with the eye or is not properly coupled. Re-aempt applicaon of the probe.
This command may be executed through voice control. Refer to the A-Scan Voice Controls secon.

Table of Contents