WARNINGS AND CAUTIONS
All warnings and cauons listed throughout this manual are to be observed when operang the
device.
WARNING: Indicates a potenally hazardous situaon which, if not avoided, could cause injury or harm to
the paent or erroneous results.
CAUTION: Indicates a potenally hazardous situaon which, if not avoided, may result in minor injury or
harm to the equipment.
GENERAL WARNINGS
WARNING: EXPLOSION HAZARD. Do not use in the presence of ammable anesthecs, gases, mixtures or in
an oxygen-rich atmosphere.
WARNING: ELECTRICAL SHOCK HAZARD. Do not open the unit. Refer servicing to qualied service
personnel.
WARNING: CONTAMINATION HAZARD. Probes used must be cleaned and disinfected aer each use. Follow
the manufacturer’s instrucons when using disinfectants.
WARNING: BURN HAZARD: The Scanmate Flex does not provide protecve means against burns to the
paent if used with HF surgical equipment. The Scanmate Flex should not be used with HF surgical
equipment.
WARNING: If the Scanmate Flex is used with other devices, current leakage may increase and electric shock
may be caused. It is the user’s responsibility to ensure safety when the probe is to be used with other
devices. If safety cannot be ensured, use of the probe with other devices is not allowed.
WARNING: The device should only be powered using the provided AC adapter. Likewise, the computer
hosng the device should only be powered using a medical grade AC/DC power supply. Use of other
adapters could potenally cause harm to the system, the probe, the operator and/or the paent.
WARNING: Device contains a Lithium Ion baery. Improper service/replacement of the baery may cause
damage to the device, excessive temperatures, re or explosion. The baery is only to be serviced/replaced
by qualied service personnel.
WARNING: Using non-essenal soware in conjuncon with the system could have unknown/adverse
impact on the operaon of the device and is therefore not recommended.
WARNING: Due to the threat of computer viruses, it is recommended that an an-virus program be installed
on the computer running the Scanmate applicaon and that paent records be backed up regularly.
WARNING: Electromagnec interference may cause distoron of received ultrasonic signals. Such
interference may result in distorted images or waveforms.
WARNING: Use of this equipment adjacent to or stacked with other equipment may result in improper
operaon and should be avoided. If such use is necessary, all equipment should be observed to verify normal
operaon.
WARNING: Portable RF communicaons equipment (including peripherals such as antenna cables and
external antennas) should be used no closer than 30 cm (12 inches) to any part of the Scanmate Flex,
including cables specied by the manufacturer. Otherwise, degradaon of the performance of this
equipment could result.
WARNING: Use of accessories, transducers and cables other than those specied or provided by the
manufacturer of this equipment could result in increased electromagnec emissions or decreased
electromagnec immunity of this equipment and result in improper operaon.
WARNING: Probes and cables should not be autoclaved or subjected to intense heat.
GENERAL CAUTIONS
CAUTION: Do not allow sharp objects, such as scalpels or cauterizing knives, to touch the probe or cables.
CAUTION: Disconnect the Scanmate Flex from the host computer before performing roune maintenance or
cleaning. Always follow the manufacturer’s instrucons when cleaning and disinfecng probes.