43
6. Aer a large populaon of results has been acquired for a given doctor and IOL model,
navigate to the Personalized Lens Constants Table to view the Procedures that were
performed by the selected doctor using the selected lens. Review the Procedures that have
Use Recordselected and modify as necessary. See step 5 for consideraons on ltering
which results to use.
7. The New Personalized Constantsgroup box shows the results of the Personalized Lens
Constants calculaon based on the Procedures that are currently selected in the table. The
Current Personalized Constantsgroup box displays the lens constants that are currently
being used for the selected Doctor and IOL.
8. Click the [↑] buon to replace the Current Personalized Constantswith the New
Personalized Constants”. These new Personalized Lens Constants will now be available for
use in future IOL power calculaons.
NOTE: Automacally calculang lens constants is similar to the procedure described above
except that the lens constants are recalculated as soon as Add to Personalized Constants
Table checkbox is selected.