USB INTERFACE MODULE (FLEX-USBIM)—GENERAL CLEANING
CAUTION: The USB interface housing is not water ght. Do not saturate or immerse the
housing in liquid.
The USB Interface Modules plasc housing can be cleaned using a so cloth dampened with
70% (v/v) Isopropyl Alcohol.
A-SCAN PROBE (DGH6006)—GENERAL CLEANING AND DISINFECTION
CAUTION: Do not submerge the A-Scan probe in liquid beyond the cord strain relief. Care
must be taken to prevent liquids from entering the probe housing from the cord-side or
from coming into contact with the probe connector.
NOTE: Disinfecon should be performed aer general cleaning.
NOTE: Wear protecve gloves when performing general cleaning or disinfecon.
GENERAL CLEANING: Disconnect the probe from the USB interface module. Dampen a so cloth
with 70% isopropyl alcohol (70% IPA). Wipe the probe p, housing and cord (if necessary) to
remove any body uids or parculate maer present. Wipe with a water-dampened cloth (using
dislled, sterile or de-ionized water) to remove any residue and then wipe dry with a so, dry
cloth.
DISINFECTION: Immerse approximately 1/4” of the probe p in a soluon of 70% isopropyl
alcohol (70% IPA). If it is desired that more of the probe housing be disinfected, the probe may
be immersed as far as the beginning of the probe cord strain relief . Allow the probe to soak for
no more than 1 hour. Rinse using dislled, sterile or de-ionized water. Thoroughly dry the probe
using sterile gauze.
A-SCAN IMMERSION SHELL (P-102)—GENERAL CLEANING AND DISINFECTION
WARNING: The catheter tubing used with the immersion shell can not be reused.
CAUTION: The immersion shell cannot be autoclaved.
The A-Scan immersion shell is not sold sterile, and must be cleaned and disinfected before each
use. Clean and disinfect the immersion shell using FDA and EPA approved cleaning and
disinfecon products. The soluons must be compable with polycarbonate and Kynar plasc.
Follow the instrucons on the cleaning and disinfectant labels. Refer to the immersion shell
manufacturers instrucons for more informaon.