Page 32 of 87 DGH 55B-INS-OMENG Rev: 3
10.5 Clearing Patient Information
10.5.1 Press the CLR key. The user will be prompted with
the ‘What To Clear?’ menu:
• Use the or key to scroll through the clearing options. Press the ENT key to
select an option.
• The following selections affect the patient currently being measured.
o Selecting ‘All Current Meas’ clears all measurements of both the left and
right eye for the currently loaded patient.
o Selecting ‘OD Current Meas’ clears all measurements of the right eye for
the currently loaded patient.
o Selecting ‘OS Current Meas’ clears all measurements of the left eye for
the currently loaded patient.
o Selecting ‘Nothing (Exit)’ exits the menu without deleting anything.
• The other clearing options affect stored patient information or system configuration.
o Selecting ‘All Patients’ clears all identifying information and all
measurements for all patients. (Bluetooth
®
option only)
o Selecting ‘Paired Devices’ clears all paired devices from the device
memory. (Bluetooth
®
option only)