Page 48 of 87 DGH 55B-INS-OMENG Rev: 3
13.3 Parameters for Mapping Measurement Mode
These are the Mapping Measurement Mode parameters that are accessible from the
Configuration Menu. The device must be set to Mapping to access these parameters.
NOTE: To access these parameters, enter the Configuration Menu by pressing the CFG key.
Press the ENT key repeatedly to navigate through the menu. See section 13.1
13.3.1 Numb of Posn (default is 33) is the number of
positions where the device will take single
measurements. It can be adjusted from 1 to 33.
Press the or key to change the value.
WARNING: All measurements in the device, including those in memory will be cleared
when the Number of Positions is changed.
13.3.2 Disp Bias Meas (default disabled) enables or
disables if the device will display a measurement
bias with each measurement taken during Mapping
Measurement Mode. Press the or key to
change the configuration
13.3.3 Amount Of Bias (default 100%) determines the
amount of bias used in calculating the measurement
bias in Mapping Measurement Mode. This
parameter is only available if the device is
configured to display measurement bias. It can be
adjusted from 1% to 199%. Press the or key to
change the value.
• The amount of bias is applied to all patients stored in the device memory. If the
amount of bias is changed, the device will re-calculate all bias measurements for all
stored patients.
13.3.4 Good Meas Delay (default 1.0 sec) is the amount
of time that the device will wait before advancing to
the next mapping point after a measurement is
taken. It can be adjusted from 1.0 sec to 9.5 sec.
Press the or key to change the value.