Page 28 of 87 DGH 55B-INS-OMENG Rev: 3
10.2 Multi-Patient Mode
NOTE: Use of Multi-Patient Mode is only available with a Bluetooth
®
enabled device. This
includes transferring data and storing data in Non-Volatile memory.
Multi-Patient Mode allows the user to access the memory
of the device for storing patient measurements. Multi-
Patient mode is indicated by a number (P01-P20) being
displayed in the bottom left-hand corner of the screen.
The memory is capable of storing information for 20 patients. The user selects a number (P01-
P20) and takes measurements for the right and left eye. When measurement is complete, the user
presses the CFG key and all results are recorded and can be retrieved at a later time.
The user can also use the DGH Connect Software (See section 11) to enter patient identification
information before taking measurements.
NOTE: In Multi-Patient Mode the device can take either Continuous Averaging or Mapping
measurements. However, if the Measurement Mode is changed it will clear the measurements for
all patients stored in memory.